PyGC – High pressure pyrolysis (CDS HPR) with GC – FID/TCD (Agilent 7890B)
High pressure pyrolysis (up to 34 bar/500 psi) with catalyst reactor, reactive gas analysis (CO2 and steam) coupled with GC – TCD for permanent gas analysis and FID for heavier organics.
Pyrolysis:
- Model: CDS HPR
- Heating Rate: Up to 20°C/ms or 2 × 104°C/s
- Maximum temperature: Up to 1400 °C
- Maximum pressure: 500 psi (3.4 MPa)
- Gasses: He, CO2, steam
- Catalyst reactor:
- Volume: 2 ml
- Maximum temperature: Up to 800 °C
- GC:
- Model 7890B
Fixed gas analysis
- Detector: TCD
- Column: Hayesep Q, Molsieve 10X
- Calibrated gases: H2, CO, CO2, CH4, C2H4, C2H6, C3H6 (0.01-1%)
Organics analysis
- Detector: FID
- Column HP-5